Design and simulation of a short, 10 MV, variable energy linear accelerator for use in linac-MR systems
DownloadFall 2016
Hybrid linac-MR devices are currently being designed and constructed by several groups around the world which combine a linear accelerator with an MRI in order to enable real-time 3D imaging to visualize tumor movement during external beam radiotherapy treatments. Each of the current systems is...